What is the difference between mode slow and fast? is the slow mode is averaging readings? how many?
The sampling time is fixed. 10 ms per reading.
Fast: within 200 ms, for ex: take 20 readings, add them up and divide them by 20 show the answer on the display.
Slow: within 800 ms, for ex: take 80 readings, add them up divide it by 80 and show the answer on the display.
The difference between slow and fast measurements is that a slow measurement stabilizes the measurement by taking a longer period of time to take the readings and taking more measurements (because there is more time to take measurements), therefore, the average tends to provide a better result.