Program timing and range limits for dynamic sequences with 85XX DC Loads
2019-12-23
The 8500 series DC electronic loads are fully programmable and have an internal built-in feature called "list mode" that allows storing and running dynamic sequence profiles to create various test scenarios. Internally, each model in the series has a built-in 1 kHz transient generator that allows it to create transient conditions up to 1ms timing between load changes.
When using list mode, knowing the timing for programming sequences is important. Below lists the adjustable timing range as well as the programmable value range of the 8500 series loads in all available modes using List mode function.
- CC Mode - 1ms to 6000ms (per programmed step); 0.001 A to rated maximum
- CV Mode - 1ms to 6000ms (per programmed step); 0.10 V to rated maximum
- CW Mode - 2ms to 6000ms (per programmed step); 0.01 W to rated maximum
- CR Mode - 20ms to 6000ms (per programmed step); 0.10 ohm to 4000 ohms
The programmable value ranges for each mode listed above also indicate the maximum programmable resolution corresponding to each mode. Except for CC mode, which has a programmable resolution of 0.001 A, the other 3 modes (CV, CW, CR) have a programmable resolution of 0.01 (V, W, ohms for CV, CW, CR modes respectively). (Note: This applies to list mode only. under regular static operating, maximum resolution is different. See datasheet specifications for
Featured Products:
8500
300W Programmable DC Electronic Load
8502
300W Programmable DC Electronic Load
8510
600W Programmable DC Electronic Load
8512
600W Programmable DC Electronic Load
8514
1200W Programmable DC Electronic Load
8518
1200W Programmable DC Electronic Load
8520
2400W Programmable DC Electronic Load
8522
2400W Programmable DC Electronic Load
8524
5000W Programmable DC Electronic Load
8526
5000W Programmable DC Electronic Load